Pages

Popular Posts

Thursday, December 22, 2016

The Acquisition of EMG Signals.



demo


Getting EMG(Electromyography) signals from the muscles is quite a difficult task. Because the amplitude of the EMG signal is varying between 50uV and 30mV. Therefore amplifiers with higher Gain must be needed for amplification. Since the input signals are in an ultra-low voltage rage, lots of noise signals pick-up during the amplification process. So series of filters are used with cutoff frequencies 0 and 500Hz. Because the frequency range of the EMG signals is between 0Hz and 500Hz. There are several steps or stages for getting  EMG signals from the muscles.

Electrodes

Copper(Cu) is a good conductive material that can be found easily. Therefore electrodes can be made using copper plates. The signal condition of the signal depends on the shape and the area which touches with the skin. According to researches about Surface Electrodes, circular shape is better than other shapes and the optimum diameter of the electrode is 10mm.

Filters & Amplifies

Three electrodes are used for getting signals. One electrode is placed on the non-muscular part of the body and it's the reference electrode (Ground Electrode /Passive Electrode) which is connected to the ground terminal of the circuit. The other two electrodes (Active Electrodes) are placed on the muscle. (An electrode is placed on the middle point of the muscles and another one is placed on the edge of the muscles). For this experiment, the calf muscle is the target muscle.
In the beginning, high frequencies are cut off before the differential amplification stage. Also, the input impedance of the differential amplifier must be at a higher level. So an instrumentation amplifier is used as the differentials amplifier. Because it has high stable gain and high input impedance. Therefore Low Pass Passive Filters with 486Hz cutoff frequency are used as the first stage for both active electrodes inputs.
It does not decrease the input impedance. The second stage is the instrumentation amplifier stage. According to the spectrum and output waveform, it can be identified that the frequencies below the 16Hz do not affect the output signals and there was a kind of floating output. Therefore the instrumentation amplifier is configured as High Pass Filter with 16Hz cutoff frequency.
The characteristic of the above circuit is based on the frequency response. A simulated frequency response is shown below.

Simulated frequency response (Using TINA)
According to the frequency response curve, the whole circuit acts as a bandpass filter with 16Hz of lower cutoff frequency and the 500Hz with upper cutoff frequency (approximately). The prototype of the circuits is shown below and it consists of other circuits (Notch filter, Gain Amplifier, Rectification Circuit, and Pass Filters).

The instrumentation amplifier which is used in the circuit is INA128 (datasheet). It has a high CMRR (120dB). The output waveform of the circuit is shown below.
Output waveform
The EMG analysis of output for contraction and relaxation of the calf muscle can be shown below. There are time-domain wave-forms(blue) and frequency domain wave-forms(red) for both states. 50Hz supply frequency can be seen here.
Comparison